075582814553
2 MOPP inlet connectors add medical-grade protection

FREE-SKY (HK) ELECTRONICS CO.,LIMITED / 04-14 13:45

Schaffner has introduced a series of filtered IEC mains input connectors designed not to compromise two MOPP (means of patient protection) safety precautions in medical equipment.

Called FN9274, they have been designed according to Class II medical equipment standard IEC / EN 60601-1.

The enclosure is plastic for higher dielectric strength insulation and C18 inlet receptacles do not have an earth pin – the input is also physically compatible with the company’s C17 V-Lock locking power cords to avoid accidental disconnection of power.


20201013223136740364FN9274 circuit

Three mounting configurations are available: front panel mount, rear panel mount and snap-in fit.

There are also two termination styles for each mounting configuration: fast-on connectors or wire leads.

They are rated to 4kVac, with creepage/clearance of 8mm/5mm and are double insulated.

Rating is 250Vac 50/60Hz up to 15A continuous.

According to the company, they fulfill the requirements for:

  • Two MOPP in IEC 60601-1 – medical

  • IEC/EN 62368-1 – audio, video, information and communication

  • IEC/EN 60950-1 – information technology

  • IEC/EN 60335-1 – household

  • Reinforced Insulation in IEC 61010-1 – Safety requirements

  • RoHS and REACH regulation

All versions are certified according UL/IEC 60939-3, C22.2 No. 8. CQC/CCC GB/T15287 and GB/T15288

Applications foreseen include general medical devices (MDD), in-vitro diagnostic medical devices (IVDD) and intensive care, as well as equipment for test, measurement, audio, video and communication


Processed in 0.397815 Second , 24 querys.